Beyond 1-channel syringe pump BYZ-810D

Features of Single Channel Syringe Pump

1. HD LCD Display, high capacity words, friendly user interface, dynamically display working status;

2. Audible and visual alarm for occlusion, empty, near empty, low battery, end of infusion, syringe loose, wrong setting, etc.;

3. Compatible with 5ml, 10ml, 20ml, 30ml, 50/60ml syringe of any brands;

4. Preset solution volume to greatly reduce the workload of nurses;

5. Three working modes: Rate mode, Time Volume mode, Dosage Weight mode;

6. Three levels of occlusion: high, medium and low;

7. Purge and Bolus function;

8. KVO (keep-vein-open) automatically opens as injection is completed, KVO rate is 0.1-5ml/h(0.1ml/h step);

9. Freely Stackable: freely stack one syringe pump onto another to provide multiple solutions, which have a wide range of clinical applications;

10. Power Source: AC100—240V, 50/60Hz; Internal Battery, DC12V car charge;

11. One-key operation makes setup easy and simple;

12. Syringe plunger grabble detector, operable with one hand in a No-Germ Environment;

13. Automatically record the settings of last injection;

Log in to view price and purchase
Categories: ,

Description

Specification of Single Channel Syringe Pump

Unique Human Voice Alarm system The device triggered the human voice prompt and stop injection automatically when there is any malfunction, making the infusion process safer and more reliable.
Rate: 50/60ml Syringe: 0.1ml/h ~ 999.9ml/h (0.1ml/h step)
1,000ml/h ~ 1,500ml/h (1ml/h step)
30ml Syringe: 0.1ml/h ~ 900.0ml/h (0.1ml/h step)
20ml Syringe: 0.1ml/h ~ 600.0ml/h (0.1ml/h step)
10ml Syringe: 0.1ml/h ~ 300.0ml/h (0.1ml/h step)
Flow Rate Accuracy Within ±3% (after correct calibration)
Mechanical precision Within ±2%
Bolus Rate 50/60ml Syringe: 1,200ml/h
30 ml Syringe: 720ml/h
20 ml Syringe: 480ml/h
10 ml Syringe: 240ml/h
Purge Rate 50/60ml Syringe: 1,500ml/h
30ml Syringe: 900ml/h
20ml Syringe: 600ml/h
10ml Syringe: 300ml/h
Volume Limit 0.1ml ~ 9999.9ml (0.1ml step)
Total Injection
Volume
0.1ml ~ 9999.9ml (0.1ml step)
Occlusion of Single Channel Syringe Pump High: 800mmHg ±200mmHg (106.7kPa±26.7kPa)
Medium: 50/600mmHg ±100mmHg (66.7kPa±13.3kPa)
Low: 300mmHg ±100mmHg (40.7kPa±13.3kPa )
Alarms Injection soon finish, end of injection,occlusion, improperly installation of syringe, wrong setting, low battery, syringe loose etc.
Power Source AC 100V ~ 240V, 50/60Hz; Internal rechargeable Li battery, capacity≥1,600mAh, 4 hours internal battery backup
KVO rate 1ml/h
Fuse F1AL/250/60V, 2pcs inside
Power Consumption 30VA
IP Classification IPX4
Equipment Classification Class II, internal power supply, Type CF
Operating Condition Ambient temperature: +5℃~ +40℃
Relative humidity 20 ~ 90%
Transport & Storage Condition Ambient temperature:-30℃~ +55℃
Relative humidity ≤95%
Dimension 280mm (L) ×210mm(W) ×130mm(H)
Weight 2.2kg (net weight)